equipment:vacuum:gp:igc
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| equipment:vacuum:gp:igc [2013/08/17 23:13] – [307] mcmaster | equipment:vacuum:gp:igc [2025/08/04 21:23] (current) – external edit 127.0.0.1 | ||
|---|---|---|---|
| Line 23: | Line 23: | ||
| {{: | {{: | ||
| - | The DA-15 male-male connector between the two units is straight through. | + | The DA-15 male-male connector between the two units is straight through. |
| + | |||
| + | "Degas can not be activated unless the IG pressure is below 5 x 10-5 in torr/mbar units, or 6.6 x 10-3 in pascal units." | ||
| + | |||
| ===== Power supply ===== | ===== Power supply ===== | ||
| Line 47: | Line 51: | ||
| | 1 | Anode grid | | | | 1 | Anode grid | | | ||
| | 2 | Filament | | | | 2 | Filament | | | ||
| - | | 3 (raised) | Do not connect | + | | 3 (raised) | N/C | Structural: supports top of anode grid. Electrically disconnected at STP | |
| | 4 | Filament | | | | 4 | Filament | | | ||
| | 5 | Anode grid | | | | 5 | Anode grid | | | ||
| Line 56: | Line 60: | ||
| {{: | {{: | ||
| + | |||
| + | |||
| + | ===== Convectron tubes ===== | ||
| + | |||
| + | Looks like its designed for GP-275 tubes although the manual just says " | ||
| + | |||
| + | Seems that Duniway Stockroom Corp. CVT-275-101 is a third party alternative. | ||
| + | |||
| ====== References ====== | ====== References ====== | ||
equipment/vacuum/gp/igc.1376781233.txt.gz · Last modified: 2013/10/20 14:59 (external edit)